86 9650-0801-01 Rev. A
Bio-Tek Instruments BP Pump
Adult/Pediatric Mode
Neonate Mode
Bio-Tek Instruments BP Pump2
Adult/Pediatric Mode
Neonate Mode
FastBP StepBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
Expected
Systolic
Expected
Diastolic
200/150 198 151 200 149
120/80 120 83 116 84
80/50 83 54 77 53
NeoBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
60/30 56 26
35/15 31 12
FastBP StepBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
Expected
Systolic
Expected
Diastolic
200/150 202 157 201 146
120/80 118 80 116 82
80/50 86 51 77 55
NeoBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
80/50 78 46
60/30 57 31
35/15 31 12