Propaq® MD Service Manual9650-0801-01 Rev. ASeptember 2010
CHAPTER 1MAINTENANCE TESTS6 9650-0801-01 Rev. AThis chapter describes the following maintenance tests:Calibration and Calibration Checks• NIBP Calibra
Calibration and Calibration Checks9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 7Calibration and Calibration ChecksNIBP Calibration CheckEquipment
CHAPTER 1MAINTENANCE TESTS8 9650-0801-01 Rev. AAcceptable ResultsThe unit passes the test if Pd-Primary is less than or equal to 3.0 mmHg and Pd-Safet
Calibration and Calibration Checks9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 9CO2 Calibration Procedure (required for annual calibration)Equipm
CHAPTER 1MAINTENANCE TESTS10 9650-0801-01 Rev. ACO2 Calibration Check ProcedureEquipment 5% CO2 calibration gas cylinder with CO2 sample line (kit ava
NIBP Valve, Pump, and Tubing Integrity Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 11NIBP Valve, Pump, and Tubing Integrity TestsNIBP Leak
CHAPTER 1MAINTENANCE TESTS12 9650-0801-01 Rev. ACorrective ActionIf the unit fails the test, the most likely cause is damaged or disconnected internal
NIBP Valve, Pump, and Tubing Integrity Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 13 NIBP Inflation TestEquipment Single lumen BP hose, a
CHAPTER 1MAINTENANCE TESTS14 9650-0801-01 Rev. AOperational TestsThe Propaq MD comes equipped with routine operational tests that can be performed per
Operational Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 15Accessing the Operational TestsThe operational tests on the unit can be accessed
ZOLL is a registered trademark, and SurePower is a trademark of ZOLL Medical Corporation. Propaq is a registered trademark of Welch Allyn, Inc. Nellco
CHAPTER 1MAINTENANCE TESTS16 9650-0801-01 Rev. AKeypad TestThis test checks all the front panel buttons on the unit to make sure they are working corr
Operational Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 17LCD TestThis test checks the colors of the Liquid Crystal Display on the unit.Fol
CHAPTER 1MAINTENANCE TESTS18 9650-0801-01 Rev. AAudio TestThis test checks the audio tones on the unit to make sure they are working correctly.Follow
Operational Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 19Battery TestThis test checks the battery to make sure it is working correctly.Fol
CHAPTER 1MAINTENANCE TESTS20 9650-0801-01 Rev. ANIBP TestThis allows the operator to perform various checks of the NIBP module to make sure it is work
Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 21Functional Verification TestsBefore You Begin the Functional Verifica
CHAPTER 1MAINTENANCE TESTS22 9650-0801-01 Rev. A Leads TestTools Needed 3-lead, 5-lead, and 12-lead cablesTest each cable separately.Test Setup Connec
Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 23Heart Rate Display TestTools Needed ECG simulatorECG Cable (3-lead, 5
CHAPTER 1MAINTENANCE TESTS24 9650-0801-01 Rev. AHeart Rate Alarm TestTools Needed ECG simulatorDo this... Observe this... Pass/Fail1.Press the power b
Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 2510. On the Propaq MD, press the Alarm Suspend key ( ).Alarm is enable
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual iTable of ContentsPrefaceOverview...
CHAPTER 1MAINTENANCE TESTS26 9650-0801-01 Rev. ALeakage Current TestTools Needed See the manufacturer’s instructions or supplied specifications for th
Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 27Shock TestTools Needed Fluke Impulse 7000 DP Analyzer, or equivalentT
CHAPTER 1MAINTENANCE TESTS28 9650-0801-01 Rev. ASynchronized Cardioversion TestTools Needed Fluke Impulse 7000DP or equivalent defibrillator analyzer.
Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 29Pacer TestTools Needed Fluke Impulse 7000 DP Analyzer, or equivalentT
CHAPTER 1MAINTENANCE TESTS30 9650-0801-01 Rev. APrinter TestTools Needed NoneTest Setup NoneDo this... Observe this... Pass/Fail1.On the Propaq MD, in
Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 31SpO2 TestTools Needed SpO2 Simulator: Fluke Index 2 or equivalent, Ne
CHAPTER 1MAINTENANCE TESTS32 9650-0801-01 Rev. AEtCO2 TestTools Needed CO2 sampling line airway adapterTest Setup NoneDo this... Observe this... Pass/
Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 33IBP TestTools Needed IBP patient simulator (with accuracy better than
CHAPTER 1MAINTENANCE TESTS34 9650-0801-01 Rev. ANIBP TestTools Needed NIBP simulator, single or dual lumen hose, adult, NIBP cuff, rigid cylinderTest
Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 35Temperature TestTools Needed IBP Temp Simulator, or equivalentTest Se
TABLE OF CONTENTSii 9650-0801-01 Rev. AChapter 4 Life Cycle CountersOverview...
CHAPTER 1MAINTENANCE TESTS36 9650-0801-01 Rev. A
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 37Chapter 2TroubleshootingOverviewThis chapter contains a list of error messages that users may see
CHAPTER 2TROUBLESHOOTING38 9650-0801-01 Rev. AQUICK ACCESS SW 1 SHORTQuick access key 1 is stuck. Replace the front panel.QUICK ACCESS SW 2 SHORTQuick
Propaq MD Error Messages9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 39PS DEFIB CHARGE SHORTPaddle set keys are stuck. Replace the paddle set or
CHAPTER 2TROUBLESHOOTING40 9650-0801-01 Rev. AIBP DEVICE FAILURE IBP software detected a problem with the IBP data that is most likely a hardware issu
Propaq MD Error Messages9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 41NIBP COMMUNICATION FAILUREThere is a communication problem with the NIBP m
CHAPTER 2TROUBLESHOOTING42 9650-0801-01 Rev. ANIBP AIR LEAK Major air leak or could not reach target pressure.Attempt to identify problems with the ho
Propaq MD Error Messages9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 43NIBP NEONATAL DETECTEDDevice thinks a neonate cuff is connected. If no neo
CHAPTER 2TROUBLESHOOTING44 9650-0801-01 Rev. ACO2 GAS INPUT OCCLUSIONCO2 inlet blocked.Replace FilterLine®. If problem persists, check for CO2 inlet o
Propaq MD Error Messages9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 45PS DEFIB ENERGY DOWN SHORTPaddle set energy down select button is stuck.Re
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 1PrefaceOverviewZOLL Medical Corporation’s Propaq MD Service Manual is intended for the trained biom
CHAPTER 2TROUBLESHOOTING46 9650-0801-01 Rev. A
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 47Chapter 3Disassembly ProceduresOverviewThis chapter provides instructions on how to disassemble an
CHAPTER 3DISASSEMBLY PROCEDURES48 9650-0801-01 Rev. A• Kapton tape• 3M copper adhesive tape, or equivalent• 1/2” nut driver• Large diagonal cutters• S
Removing the Handle9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 49Removing the HandleTools Required• #1 Phillips screwdriver with torque capabili
50 9650-0801-01 Rev. ARemoving the Printer ModuleTools Required• #1 Phillips screwdriver with torque capability• Orange (wooden) stickSetup• This proc
Removing the Printer Module9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 516 Disconnect the two (2) cable connectors from the printer module.conne
52 9650-0801-01 Rev. ADisassembling the Main UnitTool Required• # 1 Phillips screwdriver with torque capability• orange (wooden) stickSetup• This proc
Disassembling the Main Unit9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 536 Disconnect the battery communication cable (yellow, orange, brown, re
54 9650-0801-01 Rev. A11 Separate the two parts of the unit.
Removing the Front Enclosure Module9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 55Removing the Front Enclosure ModuleTools Required• #2 Phillips
PREFACE2 9650-0801-01 Rev. AChapter 6—Functional Description provides technical descriptions for the Propaq MD major subassembly modules.Appendix A—Pr
56 9650-0801-01 Rev. A6 Remove the ZIF (Zero Insertion Force) flex cable using an orange stick.
Removing the Display Module9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 57Removing the Display ModuleTools Required• #1 Phillips screwdriver with
58 9650-0801-01 Rev. A5 Flip the unit back over so the display screen is facing up.6 Lift the casting up while pushing down on the rubber boot that ru
Removing the EtCO2 and NIBP Modules9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 59Removing the EtCO2 and NIBP ModulesTool Required• # 1 Phillips
60 9650-0801-01 Rev. A3 Disconnect multi-wire IBP cable with ferrite in the area illustrated.4 Cut the tie wrap to the IBP cable in the area shown wit
Removing the EtCO2 and NIBP Modules9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 619 Remove five (5) screws and washers from the NIBP unit to rele
62 9650-0801-01 Rev. A14 Tilt side panel to access the EtCO2 connector and use an orange stick release the tabs and remove the EtCO2 connector.15 Remo
Removing the Monitor Board and SpO2 Modules9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 63Removing the Monitor Board and SpO2 ModulesTool Require
64 9650-0801-01 Rev. A3 Lift the monitor board to expose the multi-wire LCD cable on the back side of monitor board and disconnect it.4 Remove three (
Removing the Defibrillator Board and Capacitor Bank Modules9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 65Removing the Defibrillator Board and Ca
Conventions9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 3Conventions WARNING! Warning statements describe conditions or actions that can result i
66 9650-0801-01 Rev. A5 Remove screw in the middle of the plastic isolator (located on top of the processor board) in the area illustrated.Note: Durin
Removing the Defibrillator Board and Capacitor Bank Modules9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 6711 Locate the black ferrite. Using the
68 9650-0801-01 Rev. A
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 69Chapter 4Life Cycle CountersOverviewThis chapter contains instructions for setting life cycle coun
CHAPTER 4LIFE CYCLE COUNTERS70 9650-0801-01 Rev. ACO2On the Propaq MD, do the following:1. Press the Setup quick access key ( ).2. Use the navigation
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 71Chapter 5 Replacement PartsOverviewThis section contains a listing of the replacement parts availa
CHAPTER 5REPLACEMENT PARTS72 9650-0801-01 Rev. AReplacement PartsPART NUMBER QUANTITY DESCRIPTION705519 1 Handle Top705517 1 Handle Bottom705510 4 Scr
Replacement Parts9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 737772-0001-01 1 Printer Assembly7772-0002-01 1 Front Enclosure Assembly7772-0003-0
CHAPTER 5REPLACEMENT PARTS74 9650-0801-01 Rev. A
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 75Chapter 6Functional DescriptionOverviewThis chapter contains functional descriptions of the compon
PREFACE4 9650-0801-01 Rev. ATechnical Service for International CustomersInternational customers should return the unit in its original container to t
CHAPTER 6FUNCTIONAL DESCRIPTION76 9650-0801-01 Rev. AMain Processor (MP)The MP section controls the user interface, connectivity needs, as well as dat
Overview9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 77isolated serial bus through the FPGA. The IBP probes connect to the monitor board throug
CHAPTER 6FUNCTIONAL DESCRIPTION78 9650-0801-01 Rev. Amodule contains power supplies, charging and discharge circuitry, pacing circuitry, and the Pads/
Overview9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 79PrinterPropaq MD uses an internal thermal printer of 50 mm paper size that is built into t
CHAPTER 6FUNCTIONAL DESCRIPTION80 9650-0801-01 Rev. A
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 81Appendix AOverviewThis appendix includes the following diagrams:• Propaq MD Electrical Hardware In
82 9650-0801-01 Rev. APropaq MD Electrical Hardware Interconnect Diagram
Propaq MD Signal and Power Interconnect Schematic9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 83Propaq MD Signal and Power Interconnect Schematic
84 9650-0801-01 Rev. A
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 85Appendix BOverviewThis appendix includes simulators and settings that may be used to assess the pe
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 5Chapter 1 Maintenance TestsPropaq MD OverviewThis chapter includes step-by-step procedures for the
86 9650-0801-01 Rev. ABio-Tek Instruments BP PumpAdult/Pediatric ModeNeonate ModeBio-Tek Instruments BP Pump2Adult/Pediatric ModeNeonate ModeFastBP St
Overview9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 87DNI Nevada CuffLinkAdult/Pediatric ModeClinical Dynamics Smart ArmAdult/Pediatric ModeNeon
88 9650-0801-01 Rev. A