62 9650-0801-01 Rev. A
14 Tilt side panel to access the EtCO
2
connector and use an orange stick release
the tabs and remove the EtCO
2
connector.
15 Remove the EtCO2 module from the unit.
16 Remove three (3) screws from the outside
of the left side panel in the areas illustrated.
Note: During reassembly, torque screws to
6 in-lbs.
17 Remove four (4) screws from the inside top
of the left side panel in the areas illustrated
and carefully pull out side panel.
Note: During reassembly, torque screws to
6 in-lbs.
18 Remove ECG ZIF connector in the area
illustrated.
19 Remove the left side panel from the unit.
EtCo
2
connector
before removal
EtCo
2
connector
after removal
EtCO2 connector after removal