34 Multi-Function Simulators
ECG artifact
Type 50 Hz, 60 Hz, muscular, baseline wander, respiration
Size 25 %, 50 %, 100 % of the normal sinus R-Wave for each lead
Lead select All, RA, LL, LA, V1, V2, V3, V4, V5, V6
Fetal/Maternal ECG
Fetal heart rate (fixed) 60 BPM to 240 BPM in 1 BPM steps
Fetal heart rate (IUP) 140 BPM at beginning, then varies with pressure
Intrauterine-pressure
waveforms
Early deceleration, late deceleration, and acceleration
Wave duration 90 seconds, bell-shaped pressure curve, from 0 mmHg to 90 mmHg and returning to 0
IUP period 2 min, 3 min, or 5 minutes; and manual
Default settings FHR 120 BPM, early deceleration wave, manual
Invasive blood pressure
Channels 2, each independently settable with identical parameters and are individually electrically isolated from all
other signals
Input/output impedance 300 Ω ± 10 %
Exciter input range 2 to 16 V peak
Exciter-input frequency range DC to 5000 Hz
Transducer sensitivity 5 (default) or 40 µV/V/mmHg
Pressure accuracy ± (1 % of setting + 1 mmHg) accuracy guaranteed for dc excitation only
Static pressure - 10 to + 300 mmHg in 1 mmHg steps
Pressure units mmHg or Kpa
Dynamic waveforms Types (default pressures Arterial (120/80)
Radial artery (120/80)
Left ventricle (120/00)
Right ventricle (25/00)
Pulmonary artery (25/10)
Pulmonary-artery wedge (10/2)
Right atrium (central venous or CVP) (15/10)
Pressure variability Systolic and diastolic pressures are independently
variable in 1 mmHg steps
Swan-Ganz sequence Right atrium, right ventrical (RV), pulmonary artery (PA), pulmonary artery wedge (PAW)
Cardiac catheterization Chambers Aortic, pulmonary valve, and mitral valve
Respiration artifact Arterial, radial artery, and left ventricle 5 % to 10 % multiplication
Other 5 mmHg or 10 mmHg
BP output Circular DIN 5-Pin
Power-on default 0 mmHg
Respiration
Rate 0 (OFF), 15 BrPM to 150 BrPM in 1 BrPM steps
Waves Normal or ventilated
Ratio (inspiration:expiration) Normal 1:1, 1:2, 1:3, 1:4, 1:5
Ventilated 1:1
Impedance variations (Δ Ω) 0.00 Ω to 1.00 Ω in 0.05 Ω steps and 1 Ω to 5 Ω in 0.25 Ω steps
Accuracy delta ± (3 % of setting + 0.05 Ω)
Baseline 500 Ω, 1000 Ω (default), 1500 Ω, 2000 Ω, Leads I, II, III
Accuracy baseline ± 5 %
Respiration lead LA or LL (default)
Apnea selection 12 sec, 22 sec, or 32 seconds (one-time events), or continuous (Apnea ON = respiration OFF)
Power-on default 20 BrPM, delta 1.0 Ω
Temperature
Temperature 30 °C to 42.0 °C in 0.5 °C steps
Accuracy ± 0.4 °C
Compatibility Yellow Springs, Inc. (YSI) Series 400 and 700
Output Circular DIN 4-Pin
ProSim 8
Vital Signs Simulator